Sodium Analyzer Aqualyzer® 9033
AQUALYZER® 9001 SERIES
| FEATURES |
|
BENEFITS |
| Small size |
|
Low reagent and standard consumption |
| Wide analysis range: 0.01ppb – 10ppm |
|
Minimal maintenance required |
| Multi-stream (up to eight streams) |
|
Automatic operation |
| True fully automatic single or 2-point calibration |
|
|
| Low maintenance: no change of standards for up to two months (one month for reagent) |
|
|
| Missing sample flow indication |
|
|
| Quick connect electrodes |
|
|
| Probe positioning clamps |
|
|
| Automatic reference electrode electrolyte refill system (optional) |
|
|
| Grab sample capability |
|
|
| Process calibration – calibrate to sample |
|
|
| Two 4-20 mA isolated current outputs |
|
|
| Protection of electrodes from “Hot Sample” |
|
|

Aqualyzer® 9033 Sensor Unit
The sample flows through the flow switch, into the heat exchanger, passes through the constant head unit (used to stabilize minor changes in flow rate) and through the three solenoid valves. A vapor entrainment mechanism adds alkaline buffer vapor to the sample before it enters the sodium measuring flow cell and flows to drain. Calibration and grab sample are controlled automatically by the analyzer. Standard/grab sample liquid is brought into system via the three 3-way solenoid valves. A thermistor is housed in the flowcell and is used to detect the temperature of the sample/standard/grab sample. |
|
| Performance |
| Range |
|
0.01ppb - 10ppm |
| Accuracy |
|
± 5% of reading or ± 0.01ppb to (whichever is greater) within ± 5°C of calibration temperature |
| Reproducibility |
|
± 5% of reading or ± 0.01ppb (whichever is greater) at constant temperature |
| Response
Time |
|
90% of 10 – 1ppb step: less than 4 minutes; 90% of 100 – 1ppb step: less than 6 minutes |
| Temperature
Compensation |
|
Automatic |
pH Compensation |
|
Alkaline vapor buffer: 0.88 SG (35% W/V) ammonia above 0.3 ppb sodium, 50% diethylamine solution below 0.3 ppb sodium. Analyzer drain allows for easy sample pH measurement. |
| Single or 2-point Calibration |
|
Manual and fully automatic calibration. Manual initiated by operator, automatic with programmable frequency in days. No manual changes of standards during calibration. |
| Current Outputs |
|
Two isolated 4-20mA current (analog) outputs |
| Digital Outputs |
|
Communication interface via RS-232 & RS-485 using MODBUS RTU protocol |
| Alarms |
|
Missing sample alarm and three additional voltage-free contacts, alarm points set from transmitter rated at 2A, 250VAC (non-inductive) |
| Power |
|
Wide range of input power supply 90VAC – 250VAC |
| CE Certification (on request) |
|
Meets low voltage and low electromagnetic compatibility directives |
Operating
Conditions |
| Sample |
|
Temperature: 41 – 131 °F (5 - 55°C); Flow 150 – 400 ml/min |
| Ambient
Temperature |
|
32 – 131°F (0 – 55°C) |
| Pressure |
|
5 – 30 psig (0.35 – 2.07 bar) |
| Composition |
|
Sample should be filtered to 60 microns, free of film forming compounds |
Mechanical |
| Construction |
|
Transmitter – high strength ABS, IP65 (NEMA 4x) Wet Section – stainless steel sheet metal, all wetted parts stainless steel or PVC |
| Dimensions |
|
Height=33.5”(85.1 cm),Width=14.50”(36.83 cm), Depth=11.0”(27.94 cm) |
| Mounting |
|
Two key-style holes at top: Height=1.38”(3.51 cm), Top Width=0.48”(1.22 cm), Radius=0.47”(1.19 cm) Two bolt notches at bottom: Height=0.88”(2.24 cm),Width=0.48”(1.22 cm), Horizontal=10.25”(26 cm) |
| Weight |
|
33 lbs. (15 kg) |
| Wiring |
|
Eight (8) conduit knockouts. Quick connect electrodes. |
| Piping |
|
Sample Inlet: 1/4"(0.635 cm) OD Swagelok, Drain: 3/8"(0.953 cm) OD tube |
| Distance |
|
Up to 325 feet (100 meters) on request |
| Finish |
|
Corrosion resistant |